The transmitter is congured from a standard IBM compatible PC
by using the IPRO software (version 4.14 or later). IPRO 4.14 is com-
patible with Windows 98SE, Windows NT Workstation 4.0, Windows
2000 and Windows XP. When the transmitter is recongurated by a
PC no recalibration is needed.
Connect the IPRO communication cable between the serial port of
a PC (with IPRO software installed) and the PC connection port on
IPAQ-L
PLUS
.
The IPAQ-L
PLUS
transmitter do not have to be externally powered
when you congure with the IPRO+ communication cable. If you use
the IPRO-X communication cable you need to power the transmitter.
If IPRO is not installed in your PC, install the software following the
instructions in the setup program.
Conguration from PC is “on-line”, that is, the transmitter can be
installed in the process with the input connected during congura-
tion. The output is frozen while transmission from PC to IPAQ takes
place. When the transfer is done, the transmitter uses the new
parameters.
INTROD UCT ION
IPAQ-L
PLUS
is a loop-powered intelligent precision railmounted
temperature transmitter. The size and versatillity makes it an ideal
choice for all industrial instrumentation. The transmitter gives you the
possibility to use a customized linearization using 40 points, or 8 third
degree polynimials to give you the best possible precision.
IPAQ-L
PLUS
also have SmartSense, an intelligent
way to monitor the isolation resistance.
DIM ENS ION S
Intelligent Precision
Temperature
Transmitter
CONFIG URA TIO N
Since IPAQ-L
PLUS
is a precision instrument it´s recommended check-
ing the calibration once a year. Use equipment with an
inaccuracy of at least 5 times better than the checked one, e.g. if you
are checking a transmitter with an inaccuracy of 0.05% use instrument
with an inaccuracy of maximum 0.01%.
You can either perform a full recalibration of the transmitters stored
calibration parameters (Basic Calibration) or make a System
Error Correction (recommended). In each case use the IPRO software
and follow the instructions on the screen. When performing a System
Error Correction, please note that this is only a compensation for er-
rors at the measuring range in use.
CAL IBR ATI ON
86BIQ00013 2010-11
1
2
4
3
5
6
7
PC connection
ME AS URE O F SUC CE SS
USER INSTRUCTIONS
The user instruction must be read prior to adjustment and/or installation.
All information subject to change without notice.
ME AS URE O F SUC CE SS
This product should not be mixed with other kind of scrap, after usage.
It should be handled as an electronic/electric device.
www.inor.com, www.inor.se
INOR Process AB, PO Box 9125, SE-200 39 Malmö, Sweden,
Phone: +46-40-31 25 60, Fax: +46-40-31 25 70, E-mail: support@inor.se
INOR Transmitter OY, Unikkotie 13, FI-01300 Vantaa, Finland,
Phone:+358-10-421 7900, Fax: +358-10-421 7901, E-mail: support.@inor.se
INOR Transmitter GmbH, Am Bocksborn 5,
D-63571 Gelnhausen, Germany,
Phone: +49-6051 14807, Fax: +49-6051 14806, E-mail: support.de@inor.se
KROHNE Temperature Division INOR, 7 Dearborn Road,
Peabody, MA 01960, USA,
Commenti su questo manuale