The transmitter is congured from a standard IBM compatible PC
by using the IPRO software (version 4.14 or later). IPRO 4.14 is com-
patible with Windows 98SE, Windows NT Workstation 4.0, Windows
2000 and Windows XP. When the transmitter is recongurated by a
PC no recalibration is needed.
Connect the IPRO communication cable between the serial port of
a PC (with IPRO software installed) and the PC connection port on
IPAQ-L
PLUS
.
The IPAQ-L
PLUS
transmitter do not have to be externally powered
when you congure with the IPRO+ communication cable. If you use
the IPRO-X communication cable you need to power the transmitter.
If IPRO is not installed in your PC, install the software following the
instructions in the setup program.
Conguration from PC is “on-line”, that is, the transmitter can be
installed in the process with the input connected during congura-
tion. The output is frozen while transmission from PC to IPAQ takes
place. When the transfer is done, the transmitter uses the new
parameters.
INTROD UCT ION
IPAQ-L
PLUS
is a loop-powered intelligent precision railmounted
temperature transmitter. The size and versatillity makes it an ideal
choice for all industrial instrumentation. The transmitter gives you the
possibility to use a customized linearization using 40 points, or 8 third
degree polynimials to give you the best possible precision.
IPAQ-L
PLUS
also have SmartSense, an intelligent
way to monitor the isolation resistance.
DIM ENS ION S
Intelligent Precision
Temperature
Transmitter
CONFIG URA TIO N
Since IPAQ-L
PLUS
is a precision instrument it´s recommended check-
ing the calibration once a year. Use equipment with an
inaccuracy of at least 5 times better than the checked one, e.g. if you
are checking a transmitter with an inaccuracy of 0.05% use instrument
with an inaccuracy of maximum 0.01%.
You can either perform a full recalibration of the transmitters stored
calibration parameters (Basic Calibration) or make a System
Error Correction (recommended). In each case use the IPRO software
and follow the instructions on the screen. When performing a System
Error Correction, please note that this is only a compensation for er-
rors at the measuring range in use.
CAL IBR ATI ON
86BIQ00013 2010-11
1
2
4
3
5
6
7
PC connection
ME AS URE O F SUC CE SS
USER INSTRUCTIONS
The user instruction must be read prior to adjustment and/or installation.
All information subject to change without notice.
ME AS URE O F SUC CE SS
This product should not be mixed with other kind of scrap, after usage.
It should be handled as an electronic/electric device.
www.inor.com, www.inor.se
INOR Process AB, PO Box 9125, SE-200 39 Malmö, Sweden,
Phone: +46-40-31 25 60, Fax: +46-40-31 25 70, E-mail: support@inor.se
INOR Transmitter OY, Unikkotie 13, FI-01300 Vantaa, Finland,
Phone:+358-10-421 7900, Fax: +358-10-421 7901, E-mail: support.@inor.se
INOR Transmitter GmbH, Am Bocksborn 5,
D-63571 Gelnhausen, Germany,
Phone: +49-6051 14807, Fax: +49-6051 14806, E-mail: support.de@inor.se
KROHNE Temperature Division INOR, 7 Dearborn Road,
Peabody, MA 01960, USA,
Phone: +1-978-826 6900, Fax: +1-978-535 3882, E-mail: inor-inf[email protected]
Vedere la pagina 0
1 2

Sommario

Pagina 1 - Transmitter

The transmitter is congured from a standard IBM compatible PC by using the IPRO software (version 4.14 or later). IPRO 4.14 is com-patible with Wi

Pagina 2 - Input connections

LIM ITE D W ARR ANT YINOR Process AB, or any other afliated company within the Inor Group (hereinafter jointly referred to as ”Inor”), hereby warrant

Commenti su questo manuale

Nessun commento